Hemocompatibility: Hematology Platelet & Leukocyte - PLC (GLP) - Internationally Sourced

Platelet and leukocyte counts ask whether your device consumes blood cells. Human whole blood is exposed to the device and the counts afterwards are compared with a concurrent negative control.

SKU: GRK-4006-05X
Availability:
Price: Call for pricing

When this test applies

Required wherever the device contacts circulating blood, under ISO 10993-1:2025. ISO 10993-4 does not prescribe one test — it sets out which blood-interaction categories apply to which device type, and platelet and leukocyte response is one of them. It is the most direct read on whether a surface is pulling platelets out of circulation, which is the mechanism behind device-associated thrombocytopenia.

When to run it

After cytotoxicity, and once the material set and surface finish are settled. Blood interaction is sensitive to surface topography and to processing residues, so running it on a material you still intend to change tends to buy a result you have to repeat.

What a pass tells you

That neither platelet nor leukocyte counts show a significant reduction attributable to the device, measured against the concurrent negative control.

What it does not tell you

A stable count does not mean the platelets are unactivated — activation markers are a separate measurement. It also says nothing about hemolysis, complement or clotting time. ISO 10993-4 may call for several of the blood-interaction tests together, and a pass on one says nothing about the others.

Method

Testing per ISO 10993-4:2017. Human whole blood is exposed to the device or a representative sample, and platelet and leukocyte counts after exposure are compared with the concurrent negative control. Where a representative sample rather than the whole device is used, extraction ratios follow ISO 10993-12: 6 cm²/mL for material under 0.5 mm thick, 3 cm²/mL at 0.5 mm or thicker, or 0.1 to 0.2 g/mL where surface area is indeterminate.

Where it runs

This method is internationally sourced. The laboratory is ISO/IEC 17025 accredited and GLP certified, and the report is issued under Groenakker cover, in Groenakker’s report template or in your document template on request at no extra cost.

Usually ordered alongside

PTT for the coagulation arm and complement activation for the immune arm, plus hemolysis as the baseline blood-damage test.

Products specifications
SubcontractedYes
StandardsISO 10993-4:2017
EndpointHemocompatibility
Method detail
Test systemIn vitro - human whole blood
MethodHematology - platelet and leukocyte counts (PLC)
ExtractionDevice or extract incubated with anticoagulated whole blood. Extraction ratio per ISO 10993-12: 6 cm2/mL for material under 0.5 mm thick, 3 cm2/mL at 0.5 mm or thicker, or 0.1-0.2 g/mL where surface area is indeterminate.
Acceptance criteriaPer ISO 10993-4, platelet and leukocyte counts after exposure are compared with the concurrent negative control; the device passes when no significant reduction attributable to the device is observed.
DeliverableGLP-compliant final report with raw data, issued in Groenakker’s report template or in your document template on request at no extra cost, under Groenakker’s ISO 13485:2016 certified quality system.
Testing locationInternational (OUS)
*
*
*