Systemic Toxicity: Material Mediated Pyrogenicity, Rabbit (GLP) - Domestically Sourced

SKU: GRK-4004-01D
Availability:
Price: Call for pricing
Testing per ISO 10993-11:2017 / USP<151>. 1 extract (0.9% Sodium Chloride Solution) Sample Requirements: - 1 complete devices, 270cm² (>0.5mm thickness); - 1 complete devices, 540cm² (<0.5mm thickness); - 1 complete devices, 18 grams (powders/absorbable material only) Estimated Turnaround Time: 8 Weeks
Products specifications
Subcontracted Yes
Standards ISO 10993-11:2017 / USP<151>
Endpoint Material Mediated Pyrogenicity
Method detail
Turnaround 8-9 weeks
Test system In vivo — rabbit, intravenous extract
Method Material-Mediated Rabbit Pyrogen
Extraction 1 extract at 90 mL per extract. Extraction ratio per ISO 10993-12: 6 cm2/mL for material under 0.5 mm thick, 3 cm2/mL at 0.5 mm or thicker, or 0.1-0.2 g/mL where surface area is indeterminate.
Acceptance criteria Per USP <151>, the initial test uses 3 rabbits and the material passes when no individual rabbit shows a temperature rise of 0.5 °C or more. If any rabbit reaches 0.5 °C or more, the test is repeated with 5 additional rabbits (8 total); the material passes when no more than 3 of the 8 rabbits show an individual rise of 0.5 °C or more AND the sum of the 8 individual maximum temperature rises does not exceed 3.3 °C. (A single rabbit reaching 0.5 °C in the initial test triggers expansion to 8 rabbits, not outright failure.)
Deliverable GLP-compliant final report with raw data, issued in Groenakker’s report template or in your document template on request at no extra cost, under Groenakker’s ISO 13485:2016 certified quality system.
*
*
*