Systemic Toxicity: Acute Systemic Injection, 2 Extracts (GLP) - Internationally Sourced

SKU: GRK-4004-02X
Availability:
Price: Call for pricing
Testing per ISO 10993-11:2017. 0.9% Sodium Chloride Solution & Sesame Oil, NF extracts (Polar & Non-Polar) Sample Requirements: - 2 complete devices, 60cm² (>0.5mm thickness); - 2 complete devices, 120cm² (<0.5mm thickness); - 2 complete devices, 4 grams (powders/absorbable material only) Estimated Turnaround Time: 6 Weeks
Products specifications
Subcontracted Yes
Standards ISO 10993-11:2017
Endpoint Acute Systemic Toxicity
Method detail
Turnaround 6-7 weeks
Test system In vivo — mouse (or rat), polar and non-polar extracts
Method Acute Systemic Injection
Extraction 2 extracts at 10 mL per extract; at least 2 discrete dose portions. Extraction ratio per ISO 10993-12: 6 cm2/mL for material under 0.5 mm thick, 3 cm2/mL at 0.5 mm or thicker, or 0.1-0.2 g/mL where surface area is indeterminate.
Acceptance criteria Per ISO 10993-11, the material passes when test animals show no significantly greater biological reactivity than the controls — no mortality and no significant clinical signs or body-weight effects attributable to the extract — during the observation period.
Deliverable GLP-compliant final report with raw data, issued in Groenakker’s report template or in your document template on request at no extra cost, under Groenakker’s ISO 13485:2016 certified quality system.
*
*
*